Gw-instek GOM-802 Programming Manual Uživatelský manuál Strana 21

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 27
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 20
DC MILLI-OHM METER
PROGRAMMER MANUAL
40
STATus:OPERation:CONDition? (query only)
Function:
Return the contents of the OPERation register. The METER, however, do
not use the OPERation register to report any conditions.
Syntax:
STATus:OPERation:CONDition?
Returns:
<NR1>
Examples:
STATus:OPERation:CONDition? returns 0.
STATus:OPERation:ENABle
Function:
Set or query the enable mask that allows the masked conditions in the
event register to be reported in the summary bit. If a bit is 1 (true) in the
enable register and its associated event bit changes to 1 (true), the
associated summary bit will change to 1 (true). Even though this is a 16-
bit register, only 15 bits (bit 0 through bit 14) are used. Bit 15 always
reads 0.
Syntax
STATus:OPERation:ENABle <NR1>
STATus:OPERation:ENABle?
<NR1> is an integer from 0 to 32767.
Returns
<NR1>
Examples
STATus:OPERation:ENABle 32767 sets all 15 bits of the register to 1.
DC MILLI-OHM METER
PROGRAMMER MANUAL
41
If the STATus:OPERation:ENABle? returns 0, all 15 bits of the register
are 0.
STATus:OPERation:EVENt(query only)
Function:
Returns and clears the contents of the OPERation register.
Syntax:
STATus:OPERation:EVENt?
Returns:
<NR1>
Examples:
STATus:OPERation:EVENt? returns 0.
STATus:PRESet
Function:
Set the OPERation and QUESTionable enable registers to zeros.
Syntax:
STATus:PRESet
STATus:QUEStionable:CONDition? (query only)
Function:
Return the contents of the QUEStionable register. Reading the condition
register is non-destructive.
Syntax:
STATus:QUEStionable:CONDition?
Zobrazit stránku 20