64
:MEASure:CMEan
Description
Returns the mean voltage/current of one full
period.
Syntax
:MEASure:CMEan?
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the cyclic mean.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:CMEan?
9.480E-01
Selects Channel 1, and then measures the mean
value of the first period.
:MEASure:RMS
Description
Returns the root-mean-square voltage/current of
one or more full periods.
Syntax
:MEASure:RMS?
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the RMS value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Query
Query