REMOTE CONTROL
81
ON
Enables the power switch trip for OVP or
OCP.
Return parameter
<Boolean>
Returns power switch trip setting.
SYSTem:CONFigure:CURRent:CONTrol
Set
Query
Description
Sets or queries the CC control mode (local control
(panel), external voltage control, external
resistance control). This setting is applied only
after the unit is reset.
Syntax
Query Syntax
SYSTem:CONFigure:CURRent:CONTrol { 0 | 1 | 2 | 3 }
SYSTem:CONFigure:CURRent:CONTrol?
Parameter/Return
<NR1>
Description
0
Local (Panel) control
1
External voltage control
2
External resistance control; 10kΩ = Io max,
0kΩ = Io min.
3
External resistance control; 10kΩ = Io min,
0kΩ = Io max.
SYSTem:CONFigure:VOLTage:CONTrol
Set
Query
Description
Sets or queries the CV control mode (local control,
external voltage control, external resistance
control). This setting is applied only after the unit
is reset.
Syntax
Query Syntax
SYSTem:CONFigure:VOLTage:CONTrol { 0 | 1 | 2 | 3 }
SYSTem:CONFigure:VOLTage:CONTrol?
Parameter/Return
<NR1>
Description
0
Local (Panel) control
1
External voltage control
2
External resistance control; 10kΩ = Vo max,
0kΩ = Vo min.
3
External resistance control; 10kΩ = Vo min,
0kΩ = Vo max.