Gw-instek PEL-2000 Series User Manual Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Zařízení Gw-instek PEL-2000 Series User Manual. GW Instek PEL-2000 Series User Manual User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 268
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků

Shrnutí obsahu

Strany 1 - Programmable DC Electronic

Programmable DC Electronic Load PEL-2000 Series USER MANUAL GW INSTEK PART NO. 82EL-20040MA1 ISO-9001 CERTIFIED MANUFACTURER

Strany 2

PEL-2000 Series User Manual 10 GETTING STARTED This chapter describes the features and functions of the PEL-2002/2004, including the front and rea

Strany 3 - Table of Contents

PEL-2000 Series User Manual 100 9. Go/NoGo Set the Go/NoGo configurations, if applicable. Page 177 10. Protection Modes Configure the protectio

Strany 4

TUTORIALS 101 Programming When creating a Program Sequence or Chain, all channels are used at the same time unless programmed otherwise. Program Se

Strany 5 - SAFETY INSTRUCTIONS

PEL-2000 Series User Manual 102 12. Save Setup Save everything to the internal Setup memory. Page 206 13. Run Run the Program Sequence/Chain. Pa

Strany 6 - Safety Guidelines

TUTORIALS 103 Sequences Sequences are used to accurately simulate loads. As each Sequence is independent, Sequences are ideally suited to test mult

Strany 7

PEL-2000 Series User Manual 104 Frame Link Frame link connections are used connect up to four slave main frames to a master main frame. When usin

Strany 8

TUTORIALS 105 4. Run Run the loads. To run the loads, press the LOAD key on the master mainframe. To stop, press again. When the LOAD key is press

Strany 9 - SAFETY INSTRUCTIONS

PEL-2000 Series User Manual 106 Channel Control The Channel Control connectors on the rear panel can be used to control and monitor the status of

Strany 10 - GETTING STARTED

TUTORIALS 107 9. Monitor Use IMON and VMON to monitor the current and voltage of load outputs. Page 85 10. End To turn the load off, output an ac

Strany 11 - GETTING STARTED

PEL-2000 Series User Manual 108 General Configuration Options There are number of different options for each channel. The different options are d

Strany 12 - Main Features

OPERATION 109 OPERATION The PEL series operation is described in the chapters below. The sections are broken down into small operations. For thoro

Strany 13 - Series Overview

GETTING STARTED 11 Rear Panel Overview ... 25Front Panel Overview – Load Module ...

Strany 14 - PEL-2020 PEL-2040

PEL-2000 Series User Manual 110 Editing CR Static Parameters ... 132Select CV Mode ... 134Editing CV Par

Strany 15

OPERATION 111 Adjusting the Knob Control Type ... 188Configuring Alarm Sound ... 190Configuring Go/NoGo Alarm Sound .. 191Ad

Strany 16 - Measurement Overview

PEL-2000 Series User Manual 112 Local Mode Operation Each channel can be edited by its local load module. Depending on the configuration, local c

Strany 17 - Front Panel Overview

OPERATION 113 Selecting a Channel Background Each channel can be individually selected by using its load module panel. Changing channels on a load

Strany 18

PEL-2000 Series User Manual 114 Selecting Static/Dynamic Background Each load channel can be individually switched from Static to Dynamic using t

Strany 19

OPERATION 115 Turning on the Load Background Loads can be individually selected to be turned on using local operation. 1. Select a channel on t

Strany 20

PEL-2000 Series User Manual 116 Shorting Background The short key can be set to toggle or hold. 1. Ensure the load is off. 2. Select a chan

Strany 21

OPERATION 117 4. Press LOAD to turn on the load. Shorting 5. a. Press the SHORT key (toggle mode). Or b. Hold the SHORT key (hold mode). Displ

Strany 22 - Display Overview – Mainframe

PEL-2000 Series User Manual 118 Editing CC /CR/CV A/B Value Background The Selector Knob can be used to change the A Value and B Value (single ch

Strany 23

OPERATION 119 Mainframe Basic Operation For the Mainframe Basic Operation section, all operations refer to the knobs and keys on the main configura

Strany 24

PEL-2000 Series User Manual 12 Main Features Description The PEL-2002 and 2004 are multichannel programmable DC electronic load mainframes. The P

Strany 25 - Rear Panel Overview

PEL-2000 Series User Manual 120 Help Menu Background When any function key has been pressed or when a menu has been opened, the HELP key can be u

Strany 26

OPERATION 121 Channel Selection Background There are up to 2 channels per load module, depending on the model. The main display can be used to co

Strany 27

PEL-2000 Series User Manual 122 Select CC Mode Background The PEL 2000 series load generators operate in three different modes: Constant Current

Strany 28 - PEL -2020

OPERATION 123 Select CC Range Background Constant current mode can run in high and low range. Maximum range is dependent on the load module. En

Strany 29

PEL-2000 Series User Manual 124 Panel Operation 1. Press the F3 key until Dynamic Range mode is selected. Note Changing from static to dynam

Strany 30

OPERATION 125 Panel Operation 1. Use the Selector knob to highlight Level1. 03/02/0910 : 30 CCDynaHighConfigureConfigureModeCCDynamicCH1RangeHi

Strany 31

PEL-2000 Series User Manual 126 Select CC Static Mode Background Constant current mode can be set to dynamic or static mode. Static mode is for m

Strany 32

OPERATION 127 Panel Operation 1. Use the Selector knob to highlight A Value. 03/02/0910 : 30 CCStaticHighConfigureConfigureModeCCStaticCH1Range

Strany 33

PEL-2000 Series User Manual 128 Set to CR Mode Background The PEL 2000 series load operates in three different modes, Constant Current (CC), Cons

Strany 34

OPERATION 129 Panel Operation 1. Press the F2 (Range) key repeatedly until High or Low range is selected. RangeHighRangeLow The range will be re

Strany 35 - Installation

GETTING STARTED 13 Interface  USB  RS-232C  GPIB (optional) Series Overview The PEL-2000 series comprises of two different Mainframes: the PE

Strany 36

PEL-2000 Series User Manual 130 Panel Operation 1. Press F3 until Dynamic Range mode is selected. Note Changing from static to dynamic mode wil

Strany 37

OPERATION 131 Panel Operation 1. Use the Selector knob to highlight Level1. 03/02/0910 : 30 CRDynaLowConfigureConfigureModeCRDynamicCH1RangeL

Strany 38 - GPIB Installation

PEL-2000 Series User Manual 132 Select CR Static Mode Background Constant Resistance mode can be set to dynamic or static mode. Static mode is fo

Strany 39 - Rack Mount Installation

OPERATION 133 Panel Operation 1. Use the Selector knob to highlight A Value. 03/02/0910 : 30 CRStaticLowConfigureConfigureModeCRStaticCH1RangeL

Strany 40 - Channel Number

PEL-2000 Series User Manual 134 Select CV Mode Background The PEL 2000 series electronic load operates in three different modes, Constant Curren

Strany 41 - Power Up & Self Test

OPERATION 135 Editing CV Parameters Background Constant Voltage mode can be set to a maximum limit (Curr Limit). Using the current limit enables l

Strany 42

PEL-2000 Series User Manual 136 03/02/0910 : 30 CVStaticHighConfigureConfigureModeCVCH1SlowResponseSlowMain 2. Press the Selector knob to edit

Strany 43 - Load Connections

OPERATION 137 Select CV Response Speed Background Constant voltage mode has fast and slow response speeds. Quick current changes can induce line

Strany 44

PEL-2000 Series User Manual 138 Creating a Program Sequence Background The PEL series has a total of 12 different programs and 10 sequences to ea

Strany 45

OPERATION 139 Parameters Memory M001~M120 Run Skip-Auto-Manual On-Time 0.1 ~ 60.0 S Off-Time Off – 0.1 ~ 60.0 seconds P/F Time Off –

Strany 46 - Right channel

PEL-2000 Series User Manual 14 PEL-2020 PEL-2040DC Electronic LoadPEL-2002PEL-2040DC Electronic Lo ad70A / 80V , 350WPEL-2020DC Electronic LoadLR2

Strany 47

PEL-2000 Series User Manual 140 3. Press the selector knob to edit PROG:, then turn to select the program number. OR Use the number pad to ent

Strany 48 - Remote (Sense) Connection

OPERATION 141 Choose which Memory data will be used for the sequence. CH1 MXXX ~ CH8 MXXX. Run: Skip – Auto - Manual Choose whether to run t

Strany 49

PEL-2000 Series User Manual 142 The Pass(P)/Fail(F) Time can be set to 0.1 seconds less than the total test time. The total test time is defined

Strany 50 - Single Load Connections

OPERATION 143 Program Chains Background On the PEL 2000, there are up to 12 different programs containing 10 sequences. If 10 sequences in a Progr

Strany 51

PEL-2000 Series User Manual 144 03/02/0910 : 30 ConfigurePreviousMenuEditStartSaveRecallDefault 4. Use the selector knob to edit Start and con

Strany 52 - Parallel Load Connections

OPERATION 145 7. Repeat the above procedure to P02~P12 to complete the program chain. The Program Chain ends at the first Program (PXX) that i

Strany 53 - Multiple-output DUT

PEL-2000 Series User Manual 146 3. Press the FUNC key, Program (F1) and Active Channel (F2). Channel 1 (CH01) will be highlighted. Note CH1 ha

Strany 54 - Frame Link Connection

OPERATION 147 Previous Menu 9. Press F5 (Previous Menu) to return to the Sequence menu. Turn Program On/Off 10. Press F1 (Prog) to turn On or Of

Strany 55

PEL-2000 Series User Manual 148 As each sequence or program is completed the screen will update to display the active sequence/program. Notice th

Strany 56 - Channel Control Connection

OPERATION 149 Start03/02/0910 : 30 ConfigureExitDetail When the program has finished, the physical channels that have run will be displayed, a P

Strany 57

GETTING STARTED 15 Package Contents and Accessories The PEL-2000 electronic load generator has a number of standard and optional accessories that

Strany 58

PEL-2000 Series User Manual 150 Use the Selector knob to scroll down to view the remainder of the list if necessary. 17. Press F5 to exit at a

Strany 59 - Go/NoGo Connection

OPERATION 151 Edit Sequence Background The Sequence function can be configured to create a unique load profile to accurately simulate loads in rea

Strany 60 - OPERATING

PEL-2000 Series User Manual 152 3. Use the Selector knob to highlight Value. 4. Press the Selector knob to edit the Value, then turn to incre

Strany 61 - Operating Mode Description

OPERATION 153 Note The Point number can only be changed if more points have already been added. Save Sequence 11. Press Save (F3) to save the sequ

Strany 62

PEL-2000 Series User Manual 154 03/02/0910 : 30 CCStaticHighConfigurePreviousMenuCH1s_loop 2. Use the Selector knob to highlight Repeat. 3.

Strany 63

OPERATION 155 Channel Duration Time Settings Background Each sequence can have the timing duration data of another sequence. For example CH1’s seq

Strany 64

PEL-2000 Series User Manual 156 03/02/0910 : 30 ConfigureSeq.OffSaveDefineTRIG OUT Note CH1is used as the trigger source. CH2 has CH CONT set t

Strany 65 - Constant Resistance Mode

OPERATION 157 7. Press Save (F3) to save the settings. Run Sequence Background Like Programs, Sequences must be turned “ON” before they can be

Strany 66

PEL-2000 Series User Manual 158 03/02/0910 : 30 ConfigureSeq.OffSaveDefineTRIG OUT 2. Press Seq. (F1) to turn on the Sequences. 3. SEQ will

Strany 67

OPERATION 159 03/02/0910 : 30 Stop the load 6. Press the LOAD key again or wait for the Sequence (if not infinitely looped) to end/stop the load

Strany 68 - Constant Voltage Mode

PEL-2000 Series User Manual 16 Measurement Overview The PEL-2000 series has a number of different operating modes that are completely configurabl

Strany 69

PEL-2000 Series User Manual 160 Channel Configuration The Channel Configuration chapter describes the configuration options for individual channel

Strany 70

OPERATION 161 Accessing the Configuration Menu Background The configuration menu is used to access instrument settings and properties as well as s

Strany 71 - Run Program

PEL-2000 Series User Manual 162 Setting (OCP/OVP/OPP/UVP) Background Over Protection is used to set the voltage, current or power limit. In the

Strany 72

OPERATION 163 03/02/0910 : 30 CCDynaHighConfigurePreviousMenuProtectionGo-NoGoCH1OtherConf 1. Use the Selector knob to highlight OCP Level. 2

Strany 73

PEL-2000 Series User Manual 164 6. Use the selector knob to turn ON, OFF or CLEAR the OCP Setting. 7. Repeat steps 1-5 for : OVP Level OVP S

Strany 74 - Sequence

OPERATION 165 Protection Clear Background When any of the protection circuits have been tripped, the Protection Clear function can be used to rese

Strany 75

PEL-2000 Series User Manual 166 1. Press the F2(Other) key to enter the Other menu. 03/02/0910 : 30 CCDynaHighConfigurePreviousMenuProtection

Strany 76

OPERATION 167 Adjusting the Von Voltage and Latch Background The Von Voltage is the voltage point at which the load module will start to sink curr

Strany 77

PEL-2000 Series User Manual 168 3. Press the Selector knob to edit the selected value, then turn to increase or decrease the value. OR Use the n

Strany 78 - Parallel Dynamic Loading

OPERATION 169 1. Press the F2(Other) key to enter the Other menu. 03/02/0910 : 30 CCDynaHighConfigurePreviousMenuProtectionCH1Other Go-NoGoCon

Strany 79 - Configurations Description

GETTING STARTED 17 Front Panel Overview DC Electronic LoadPEL-2002LOCALCAL.F5F4F3F2ON /OFFSHIFTPRESETENTERF1FILEFUNCLOAD9CHAN873CLEAR210456P0 LOC

Strany 80 - Load Input Voltage

PEL-2000 Series User Manual 170 Configuring Channel Control Background When Channel Control (CH CONT) is set to External, it will disable editing

Strany 81 - Operating Configurations

OPERATION 171 3. Press the Selector knob to edit the selected setting, turn to change the setting from Panel to External. 4. Press the Select

Strany 82

PEL-2000 Series User Manual 172 Panel operation Ensure the menu is in the configuration menu. See page 160. 1. Press the F2(Other) key to ent

Strany 83

OPERATION 173 Note When a channel has been set to independent, an asterisk will be shown next to the channel number indicator in the Current Operat

Strany 84

PEL-2000 Series User Manual 174 03/02/0910 : 30 CCDynaHighConfigurePreviousMenuProtectionGo-NoGoCH1OtherSΩΩConf 2. Use the Selector knob to hi

Strany 85

OPERATION 175 Configuring Step Resolution Background The CC, CV and CR step resolution settings can be edited in the configuration menu. The minim

Strany 86 - IMON 10

PEL-2000 Series User Manual 176 2. Use the Selector knob to scroll down to highlight CCH Step. Note: CCH Step is off-screen when entering Other

Strany 87 - VMON 10

OPERATION 177 Go/NoGo Background Go/NoGo mode is used to set threshold limits. When a load is within the limit(s) it is considered to be “Go”, w

Strany 88

PEL-2000 Series User Manual 178 Choose Absolute/ Percentage limits 1. Press (F3) Go-NoGo to access the Go/NoGo menu. 2. Use the selector knob

Strany 89 - Interface and File System

OPERATION 179 Turn Go/NoGo On/Off 7. Use the Selector knob to edit Spec Test. 8. Choose ON to turn on Go/NoGo. 9. Choose OFF to turn off

Strany 90

PEL-2000 Series User Manual 18 System Keys CHAN/LOCAL is used to select the load channel. Combined with the shift key, Local is used to activate/

Strany 91

PEL-2000 Series User Manual 180 Mainframe Configuration The Mainframe Configuration chapter describes configuration settings that apply to all cha

Strany 92

OPERATION 181 Accessing System Information Background The System Information displays the mainframe and load module(s) serial numbers. Parameters

Strany 93 - *Excluding SEQ data

PEL-2000 Series User Manual 182 Accessing the Load Menu Background The PEL series is able to automatically start loading from the last program or

Strany 94 - File Format

OPERATION 183 4. Press the Selector knob or Enter to confirm selection. 5. Scroll to Auto Load On and repeat the steps. Configuring the Date

Strany 95

PEL-2000 Series User Manual 184 03/02/0910 : 30 ConfigureOtherTime SetSystemInfoLoad Interface 3. Use the Selector knob to highlight Month.

Strany 96 - TUTORIALS

OPERATION 185 Panel operation 1. Press the Shift Key then the Help key to access the Utility menu. 2. Press F5 (Other Menu). 03/02/0910 :

Strany 97 - Local loads

PEL-2000 Series User Manual 186 Parameters Brightness 50~90 50(low) 90(bright) Contrast 3~13 3(low) 13(high) Panel operation 1. Press the Shif

Strany 98

OPERATION 187 Adjusting the Frame Control Background Frame control is used to control a number of different frame linked mainframes (slaves) with

Strany 99 - Single Channel Load

PEL-2000 Series User Manual 188 5. Press the Selector knob to edit, then turn to turn Frame CONT (frame control) ON or OFF. 03/02/0910 : 30

Strany 100

OPERATION 189 versa. With the Knob type set to Old, independent operation is activated. Local operation changes do not change the setting/values on

Strany 101 - Programming

GETTING STARTED 19 Number pad Enter numerical values, or to save/recall presets (P0-P9). Number values. Preset numbers P0-P9. Decimal point a

Strany 102

PEL-2000 Series User Manual 190 Configuring Alarm Sound Background The PEL series has two different types of alarms, one located on the mainframe

Strany 103 - Sequences

OPERATION 191 4. Press the Selector knob to select the master alarm (Alarm Tone(M)), turn to edit and press to confirm selection. 5. Rep

Strany 104 - Frame Link

PEL-2000 Series User Manual 192 03/02/0910 : 30 ConfigureOtherSystem InfoLoad Interface Time Set 3. Use the Selector knob to move the cursor d

Strany 105

OPERATION 193 module display whilst “Measure” will show the actual measured value when editing the load. These settings apply to all channels. The

Strany 106 - Channel Control

PEL-2000 Series User Manual 194 5. Press the Selector Knob or Enter to confirm selection. View Language Settings Background The language setti

Strany 107

OPERATION 195 Interface Configuration (settings) The Interface Configuration chapter describes configuration settings that apply when using the PEL

Strany 108

PEL-2000 Series User Manual 196 Configuring RS232 Connection Background When using RS232 a number of parameters need to be set. These include Bau

Strany 109 - OPERATION

OPERATION 197 5. Press the Selector knob to confirm. 6. The RS232 Menu appears. 03/02/0910 : 30 OConfigureOtherSystemInfoInterface Time Set

Strany 110

PEL-2000 Series User Manual 198 Configuring the GPIB Address Background When using GPIB, an address must be specified. Parameters Address 01~30

Strany 111 - OPERATION

OPERATION 199 03/02/0910 : 30 OConfigureOtherSystemInfoInterface Time SetLoadGPIB 7. Use the selector knob to edit the GPIB address. Note The G

Strany 112 - Local Mode Operation

March 2010 This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be pho

Strany 113 - Selecting a Channel

PEL-2000 Series User Manual 20 Selector Knob Used to select operations and to increase/decrease values. When turned left or right moves the cur

Strany 114 - Selecting Static/Dynamic

PEL-2000 Series User Manual 200 03/02/0910 : 30 OConfigureOtherSystemInfoInterface Time SetLoadRS232 3. If the Interface mode is not USB, use

Strany 115 - Turning on the Load

OPERATION 201 Save / Recall Save / Recall ... 201Saving/Recalling Channels ...

Strany 116 - Shorting

PEL-2000 Series User Manual 202 Saving/Recalling Channels Background The PEL 2000 series can save data for up to 120 different channel configu

Strany 117 - Display Output View

OPERATION 203 5. Press the Selector knob to edit Memory (M001-M120) OR Use the number pad to enter a number. 03/02/0910 : 30 ConfigureS

Strany 118 - Editing CC /CR/CV A/B Value

PEL-2000 Series User Manual 204 Saving/Recalling Preset memory Background The PEL 2000 series can store up to 10 presets for each channel. The

Strany 119 - Mainframe Basic Operation

OPERATION 205 Save / Recall Current Channel Save / Recall All Channels 5. Press the Selector knob to edit Preset (P0-P9) OR Use the number pa

Strany 120 - ConfigureExit

PEL-2000 Series User Manual 206 Note The display will revert to the channel menu after recalling memory. Saving/Recalling Setup Memory Backgroun

Strany 121 - Channel Selection

OPERATION 207 5. Press the Selector knob to edit Setup Memory (1~4) OR Use the number pad to enter a number. 03/02/0910 : 30 ConfigureS

Strany 122 - Select CC Mode

PEL-2000 Series User Manual 208 Setting the Default USB Path/File Background When saving files to a USB memory stick the files will be saved into

Strany 123 - Select CC Dynamic Mode

OPERATION 209 03/02/0910 : 30 ConfigurePreviousSelect Rename DeleteNewFolder The top section (window) shows the current USB path. There are 5 op

Strany 124

GETTING STARTED 21 The shift key is used to access alternate functions assigned to select keys Inactive Active. When active the shift key can b

Strany 125

PEL-2000 Series User Manual 210 03/02/0910 : 30 ConfigurePreviousSelect Rename DeleteNewFolder The new path will be shown in upper Path box in

Strany 126 - Editing CC Static Parameters

OPERATION 211 9. When a key is highlighted use the selector knob, F1 or Enter to confirm a key entry. 10. Use F2 (Back Space) to delete any pr

Strany 127

PEL-2000 Series User Manual 212 The On Screen Keyboard appears 15. Use the Selector knob to scroll left and right through the keys. 16. W

Strany 128 - Select CR Range

OPERATION 213 22. To confirm deletion, press F4 again. 03/02/0910 : 30 ConfigurePreviousSelect Rename DeleteNewFolder Saving Setups to USB Me

Strany 129 - Select CR Dynamic Mode

PEL-2000 Series User Manual 214 3. Press the File key. 4. Press F1 repeatedly until the Media USB menu appears. 03/02/0910 : 30 Configu

Strany 130

OPERATION 215 03/02/0910 : 30 ConfigureSave RecallCCDynaHighCH1MediaUSBFileUtilityFile The screen will update to only show Setup files (*.S) tha

Strany 131

PEL-2000 Series User Manual 216 Saving/Recalling Memory Data to USB Background There are two options to save or recall Memory data to a USB flas

Strany 132 - Editing CR Static Parameters

OPERATION 217 3. Press the File key. 4. Press F1 repeatedly until the Media USB menu appears. 03/02/0910 : 30 ConfigureSave RecallCCDyna

Strany 133

PEL-2000 Series User Manual 218 03/02/0910 : 30 ConfigureSaveCCDynaHighCH1MediaUSBFileUtilityFile The screen updates to show Save Folder. Note

Strany 134 - Select CV Mode

OPERATION 219 03/02/0910 : 30 ConfigureSave RecallCCDynaHighCH1MediaUSBFileUtilityFile 13. Use the selector knob to edit Save or Recall File

Strany 135 - Editing CV Parameters

PEL-2000 Series User Manual 22 Display Overview – Mainframe Mainframe Status Panel The Mainframe Status Panel displays the status of the Mainfr

Strany 136 - Response

PEL-2000 Series User Manual 220 Saving/Recalling Presets to USB Background There are two options to save or recall Channel Presets to a USB flash

Strany 137 - Select CV Response Speed

OPERATION 221 3. Press the File key. 4. Press F1 repeatedly until the Media USB menu appears. 03/02/0910 : 30 ConfigureSave RecallCCDyna

Strany 138

PEL-2000 Series User Manual 222 03/02/0910 : 30 ConfigureSaveCCDynaHighCH1MediaUSBFileUtilityFile The screen updates to show Save Folder. Note

Strany 139 - ConfigureChain Save

OPERATION 223 03/02/0910 : 30 ConfigureSave RecallCCDynaHighCH1MediaUSBFileUtilityFile 13. Use the selector knob to edit Save File or Recall

Strany 140

PEL-2000 Series User Manual 224 Saving/Recalling Sequences to USB Background There are two options to save or recall Sequences to a USB flash dri

Strany 141

OPERATION 225 4. Press F1 repeatedly until the Media USB menu appears. 03/02/0910 : 30 ConfigureSave RecallCCDynaHighCH1MediaUSBFileUtilityFi

Strany 142

PEL-2000 Series User Manual 226 03/02/0910 : 30 ConfigureSaveCCDynaHighCH1MediaUSBFileUtilityFile The screen updates to show Save Folder. Note

Strany 143 - Program Chains

OPERATION 227 03/02/0910 : 30 ConfigureSave RecallCCDynaHighCH1MediaUSBFileUtilityFile 13. Use the selector knob to edit Save File or Recal

Strany 144 - 10 : 30

PEL-2000 Series User Manual 228 Quick Preset Recall/Save Background The PEL 2000 series mainframes have up to 10 Channel Presets (P0~P9). Quick

Strany 145 - Running a Program

OPERATION 229 Recall Current Channel Preset 7. Press the Preset key and one of the number keys. 0 = P0 1= P0 etc. Recall All Channel Presets 8.

Strany 146

GETTING STARTED 23 Load Type SEQPROGLOADLOADPROGSEQ The Load Type Icon indicates if a Sequence (SEQ) or Program (PROG) is turned on. If not then L

Strany 147 - ConfigureNextStop

PEL-2000 Series User Manual 230 Recall Preset Memory (Frame link) Background A master mainframe can command all units to recall preset memory fro

Strany 148

231 Recall Factory Defaults Background The Factory defaults can be recalled at any time. For details on the factory defaults please see the defa

Strany 149 - ConfigureExitDetail

PEL-2000 Series User Manual 232 INTERFACE The Interface chapter details the pin configuration of the RS232, GPIB, Frame Link, Channel Control and

Strany 150

INTERFACE 233 Interface Configuration Configure RS-232C interface RS-232C configuration Connector DB-9, Male Baud rate 9600 Parity None Data bi

Strany 151

PEL-2000 Series User Manual 234 Configure GPIB interface Connection Connect the GPIB cable to the rear panel port: 24-pin female connector. Pin

Strany 152

INTERFACE 235 Configure Channel Control interface Channel control configuration Connector Screwless connector. Wire Gauge 22-28 AWG (24 AWG recomm

Strany 153 - Create Sequence Loop

PEL-2000 Series User Manual 236 5 Load On Load On Input. Load on = Active low, 0-1V Load off = Active high, 4-5V. (Using the + load polarity in co

Strany 154 - Previous

INTERFACE 237 Pin7 MEM_2 Recall Setup memory 3 (All channels) Pin8 MEM_3 Recall Setup memory 4 (All channels) Pin9 Enable Enable Load (On/Of

Strany 155

PEL-2000 Series User Manual 238 Pin5 Sync._MEM_0 Sync signal, Recall Setup memory 1 (All channels) Pin6 Sync._MEM_1 Sync signal, Recall Setup

Strany 156

INTERFACE 239 Connection  Input: active low (0-1V), active high (4-5V)  Inputs (5V, 10kΩ pull-up resistor)  Open collector outputs (Load s

Strany 157 - Run Sequence

PEL-2000 Series User Manual 24 Mode CC CR CV Displays the current mode. Static / Dynamic Displays whether the channel is in Static or Dynamic m

Strany 158

PEL-2000 Series User Manual 240 Pin5 Ch3_GO/NG Pin13 Ch7_GO/NG Pin6 GND Pin14 GND Pin7 Ch4_GO/NG Pin15 Ch8_GO/NG Pin8 GO/NG_Enable Connectio

Strany 159

FAQ 241 FAQ Q1. The load voltage indicated on the load module is below expected. A1. Ensure the load leads are as short as possible, twisted and

Strany 160 - Channel Configuration

PEL-2000 Series User Manual 242 APPENDIX Fuse Replacement Step 1. Turn off the power at the wall socket and rear panel. Remove the power cord.

Strany 161

APPENDIX 243 Battery Replacement Background The onboard battery is used to power the real time clock and memory. Step 1. Remove the screws from t

Strany 162 - Setting (OCP/OVP/OPP/UVP)

PEL-2000 Series User Manual 244 Firmware Update Background The PEL-2000 firmware can be easily updated using a USB memory stick. For the latest f

Strany 163

APPENDIX 245 8. Turn the power Off and On from the front panel to reset the load generator. WARNING Do not turn the load generator off or remove

Strany 164

PEL-2000 Series User Manual 246 Range Chart PEL-2020 Low Range 100W 010203040506070809000.511.522.5VoltageCurrentPEL-2020 Low Range Chart PEL-20

Strany 165 - Setting the CC Voltage Range

APPENDIX 247 PEL-2030 Left Channel 30W 01020304050607080900123456VoltageCurrentPEL-2030 Left Channel Chart PEL-2030 Right Channel Low Range 250W 01

Strany 166

PEL-2000 Series User Manual 248 PEL-2030 Right Channel High Range 250W 01020304050607080900 5 10 15 20 25 30 35 40 45VoltageCurrentPEL-2030 Right

Strany 167 - Other Go-NoGo

APPENDIX 249 PEL-2040 Low Range 350W 0102030405060708090012345678VoltageCurrentPEL-2040 Low Range Chart PEL-2040 High Range 350W PEL-2040 High R

Strany 168 - Configuring the Short key

GETTING STARTED 25 Rear Panel Overview PEL-2004 PEL-2002 GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.NO OPERATOR SERVICEABLE COMPON

Strany 169

PEL-2000 Series User Manual 250 PEL-2041 Low Range 350W 01002003004005006000 0.2 0.4 0.6 0.8 1 1.2VoltageCurrentPEL-2041 Low Range Chart PEL-2041

Strany 170 - Configuring Channel Control

APPENDIX 251 Default Settings Menu Item CC Mode Mode: Static A/B Value: Min A Rising Slew Rate: Max Falling Slew Rate: Max CR Mode Mode: Sta

Strany 171

PEL-2000 Series User Manual 252 High: Max Low: Min FUNC- Program PROG: 01 SEQ: 01 Memory: 001 Run: Skip On-Time: 0.1s Off-Time: Off P/F-Ti

Strany 172

APPENDIX 253 Specifications PEL-2002 PEL-2004 MODULE SLOTS 2 4 GENERAL Operating Environment Temperature 0°C to 40°C Altitude Up to 2

Strany 173 - 0~10 S

PEL-2000 Series User Manual 254 PEL-2020 (100Wx2) CHANNELS 2 CHANNEL Left/Right Left/Right RANGE Low High POWER 100W 100W CURRENT 0~2A 0

Strany 174

APPENDIX 255 Current Accuracy ±0.4% F.S. MEASUREMENT VOLTAGE READBACK Range 0~16V 0~80V Resolution 0.5mV 2.5mV Accuracy ±(0.05%set + 0.

Strany 175

PEL-2000 Series User Manual 256 PEL-2030 (30W/250W) CHANNELS 2 CHANNEL Left Right Right RANGE N/A Low High POWER 30W 250W 250W CURRENT

Strany 176

APPENDIX 257 DYNAMIC MODE CONSTANT CURRENT MODE T1&T2 0.025mS ~ 10mS / Res : 1uS 10mS ~ 30S / Res : 1mS Accuracy 1uS / 1mS + 100ppm Sle

Strany 177 - Go/NoGo

PEL-2000 Series User Manual 258 Over Temperature Protection ؒ85Ԩ Rated Power Protection: Value 33W 275W Accuracy ±2%set ±2%set GENERAL S

Strany 178

APPENDIX 259 PEL-2040 PEL-2041 CHANNELS 1 1 RANGE Low High Low High POWER 350W 350W 350W 350W CURRENT 0~7A 0~70A 0~1A 0~10A VOLTAGE

Strany 179

PEL-2000 Series User Manual 26 RS232 port/ GPIB port The RS232 and GPIB port is used for remote control connections. RS-232C: DB-9 pin male GPIB

Strany 180 - Mainframe Configuration

PEL-2000 Series User Manual 260 Current 0~7A 0~70A 0~1A 0~10A Current Resolution 1mA 10mA 0.2mA 2mA Current Accuracy ±0.4% F.S. ±0.4% F.S

Strany 181 - Interface Time SetLoad

APPENDIX 261 Dimensions PEL-2002/PEL-2004

Strany 182 - Accessing the Load Menu

PEL-2000 Series User Manual 262 PEL-2020/PEL-2030/PEL-2040/PEL2041

Strany 183

APPENDIX 263 EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. (1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan (2) No.

Strany 184 - Adjusting the Speaker

PEL-2000 Series User Manual 264 INDEX Alarm configuration ... 190 All Chan save/recall description ... 93 Batter

Strany 185 - Load Interface Time Set

INDEX 265 Enter key ... 21 Environment operation ... 8 Storage ...

Strany 186

PEL-2000 Series User Manual 266 Mainframe configuration alarm sound ... 190 Brightness & contrast ... 185 Config

Strany 187 - Adjusting the Frame Control

INDEX 267 Operating Configurations – CC vrange ... 81 Operating Configurations – independent ... 83 Ope

Strany 188

PEL-2000 Series User Manual 268 Quick save Presets to internal memory ... 228 Recall factory defaults ... 231 Recal

Strany 189

GETTING STARTED 27 Power Socket/ Fuse The power supply socket accepts the AC mains Voltage. The fuse holder is located below the power socket. Po

Strany 190 - Configuring Alarm Sound

PEL-2000 Series User Manual 28 Front Panel Overview – Load Module PEL -2020DC Electronic LoadLR20A / 80V , 100W20A / 80V , 100WR/LDYNA.STATIC/SHOR

Strany 191

GETTING STARTED 29 Display Key Used to alternate the display output on the load module. Current Voltage Power Load time Load Key LOAD Activat

Strany 192

Table of Contents 3 Table of Contents SAFETY INSTRUCTIONS ... 5GETTING STARTED ...

Strany 193

PEL-2000 Series User Manual 30 Positive and Negative Terminals Righ t For 2 channel load modules, the left terminals are used for the 1st channel

Strany 194 - View Language Settings

GETTING STARTED 31 Short Key The SHORT key is used to manually short circuit the active channel on the local active load. When a load is off, the

Strany 195

PEL-2000 Series User Manual 32 LED Display Overview – Load Module 1&2. Channel Display or Left and right channel indicator. 5 digit d

Strany 196 - Configuring RS232 Connection

GETTING STARTED 33 4. Mode Indicator The Mode Indicator LEDs will indicate what the current mode or settings are on the active channel(s). or

Strany 197

PEL-2000 Series User Manual 34 Lights up when a load is shorted. Lights up when Go/NoGo is activated and the load fails (NG) the Go/NoGo limits.

Strany 198 - Configuring the GPIB Address

GETTING STARTED 35 Installation The installation chapter describes how to load the different load modules, install the optional GPIB card, the rack

Strany 199

PEL-2000 Series User Manual 36 3. Use the supplied screw to fix the module to the load slot, located under the load terminals. 4. Install

Strany 200

GETTING STARTED 37 6. Use the supplied screws to fix the panel cover(s) over the load slot.

Strany 201 - Save / Recall

PEL-2000 Series User Manual 38 GPIB Installation WARNING To avoid static electricity, please use appropriate anti-static work practices. GPIB C

Strany 202 - ConfigureSave Recall

GETTING STARTED 39 Rack Mount Installation Background The PEL-2004 can be used in a standard 19” rack mount enclosure with the optional rack mount

Strany 203

PEL-2000 Series User Manual 4 Frame Link ... 104Channel Control ... 106Gener

Strany 204

PEL-2000 Series User Manual 40 Channel Number Description The channel number for a module load is determined by which slot it occupies on the ma

Strany 205

GETTING STARTED 41 Power Up & Self Test Panel operation 1. Connect the power cord to the power socket. 2. Turn the external power switch

Strany 206

PEL-2000 Series User Manual 42 When the system check happens, the load modules will display each channel as it is checked, then display the cu

Strany 207

GETTING STARTED 43 Load Connections Precautions and Procedures Intoduction The PEL-2000 load generator supports a number of different load configu

Strany 208

PEL-2000 Series User Manual 44 Wire Selection Wires that are selected must be large enough to withstand a short circuit and limit voltage drops t

Strany 209

GETTING STARTED 45 The diagram above shows how changes in current can affect voltage. Limiting Load line inductance Load line inductance can be

Strany 210

PEL-2000 Series User Manual 46 Load module considerations The PEL-2000 supports single and dual channel load modules. Single channel load modules

Strany 211

GETTING STARTED 47 4. Close the terminal cover securely. Ensure the wires are secured properly and that the wires are not exposed when the cove

Strany 212

PEL-2000 Series User Manual 48 Remote (Sense) Connection Background The electronic load modules have two voltage sense contacts: Vsense L(black),

Strany 213 - Saving Setups to USB Memory

GETTING STARTED 49 Remote Sense Terminal connection The voltage sense terminals use a screw-less clamp connector. The clamp must be opened prior to

Strany 214

SAFETY INSTRUCTIONS 5 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow when operating the PEL-2002/PEL-2

Strany 215

PEL-2000 Series User Manual 50 Single Load Connections Dual Channel Load Module Connection A dual channel load module can be used to sink two l

Strany 216

GETTING STARTED 51 DC Connection For purely DC operation, a resistor and capacitor can be connected in parallel to the electronic load to reduce os

Strany 217

PEL-2000 Series User Manual 52 Parallel Load Connections Parallel load modules When the power output of a DUT exceeds the power rating of a chan

Strany 218

GETTING STARTED 53 Parallel loads using framelink connections The PEL-2000 mainframes can also be connected in parallel. Please note, when using a

Strany 219

PEL-2000 Series User Manual 54 Frame Link Connection Background Frame link control involves connecting multiple mainframes using the frame link

Strany 220

GETTING STARTED 55 The first mainframe that is connected is the master frame; any additional frames are slave units. The ribbon cable connects to

Strany 221

PEL-2000 Series User Manual 56 Channel Control Connection Background The Channel Control connecters are located on the rear panel of each mainf

Strany 222

GETTING STARTED 57 External Voltage Connection The external voltage reference input must be between 0~10V. WARNING Ensure the external voltage r

Strany 223

PEL-2000 Series User Manual 58 Voltage and Current Monitor Output The Voltage Monitor Output (VMON) and Current Monitor Output (IMON) output the l

Strany 224

GETTING STARTED 59 Go/NoGo Connection Background The Go/NoGo port is a 15 socket port. Each channel has a dedicated line for a Go/NoGo output. T

Strany 225

PEL-2000 Series User Manual 6 Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or conta

Strany 226

PEL-2000 Series User Manual 60 OPERATING DESCRIPTION Operating Mode Description ... 61Constant Current Mod

Strany 227

GETTING STARTED 61 Operating Mode Description There are three basic operating modes: constant current (CC), Constant Resistance (CR), and Constant

Strany 228 - Quick Preset Recall/Save

PEL-2000 Series User Manual 62 Range There are two selectable ranges for constant current mode: high and low range. Low range has a higher resolut

Strany 229

GETTING STARTED 63 Slew rate The slew rate is the rate at which the current will increase to a set level. There are two slew rates: rising slew r

Strany 230

PEL-2000 Series User Manual 64 Load CurrentVoltageHighGoLowCC Mode: Static:Go/NoGoCurrentNoGoNoGo GO is specified as between the Low and High

Strany 231 - ConfigureRecall

GETTING STARTED 65 Constant Resistance Mode Background In Constant Resistance Mode the load units will linearly sink current and voltage to matc

Strany 232 - INTERFACE

PEL-2000 Series User Manual 66 Single Load Module Dynamic Functions CR mode supports Dynamic loading. Dynamic load has two resistance levels (Le

Strany 234 - Configure GPIB interface

PEL-2000 Series User Manual 68 Constant Voltage Mode Background In Constant Voltage Mode the load units will sink current whilst keeping the vol

Strany 235

GETTING STARTED 69 CV ModeVoltageTimeA valueB value CV + CC When using CV mode, a current limit can be set for CV + CC mode. When the voltage inp

Strany 236

SAFETY INSTRUCTIONS 7 Power Supply WARNING  AC Input voltage: 115V/230V switchable, 50/60Hz  The power supply voltage should not fluctuate more

Strany 237

PEL-2000 Series User Manual 70 Response Speed Response speed can be set to fast or slow. Fast response and slow response is determined by the lo

Strany 238

GETTING STARTED 71 Run Program Background The Program function on the PEL-2000 series supports a total of 12 different programs at any one time

Strany 239 - Configure Go/NoGo Interface

PEL-2000 Series User Manual 72 P/F-Time Off – 0.1 ~ (On-time+Off-time)-0.1 Short Channel CH1 ~ CH8 Program Sequences are run sequentially to

Strany 240 - USB Interface Connection

GETTING STARTED 73 Go/NoGo Results If Go/NoGo limits have been configured, the Pass/Fail results for each channel will be displayed for all the seq

Strany 241

PEL-2000 Series User Manual 74 Sequence Background The Sequence function is used to create high resolution load simulations. Each Sequence can

Strany 242 - APPENDIX

GETTING STARTED 75 Points Up to 120 points can be used with each Sequence. Each point can have a different duration, slew rate and value. A new poi

Strany 243 - Battery Replacement

PEL-2000 Series User Manual 76 On end load If more than one Sequence is programmed on the mainframe, the On End Load function allows a Sequence to

Strany 244 - Firmware Update

GETTING STARTED 77 Channel Duration Time Setting The Channel Duration Time Setting feature allows the point time duration of one Sequence to be imp

Strany 245 - Calibration

PEL-2000 Series User Manual 78 Parallel Dynamic Loading Background The PEL-2000 series of DC electronic loads support parallel dynamic loading.

Strany 246 - Range Chart

GETTING STARTED 79 Configurations Description There are a number of different configurations for the PEL series including protection modes, operati

Strany 247 - PEL-2030 Left Channel Chart

PEL-2000 Series User Manual 8 Operation Environment  Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution (Note belo

Strany 248

PEL-2000 Series User Manual 80 Over Current Protection When a load unit is operating in CR or CV mode, the unit may need over current protectio

Strany 249 - PEL-2040 High Range Chart

GETTING STARTED 81 Reverse voltage Protection Reverse voltage protection prevents reverse voltage damage to the PEL-2000 series up to the specifi

Strany 250 - PEL-2041 High Range Chart

PEL-2000 Series User Manual 82 Von Voltage Von Voltage is the voltage limit at which the load will start to sink current. There are two operatio

Strany 251 - Default Settings

GETTING STARTED 83 Short When short mode is on, the load unit can simulate a short circuit. Shorting can be individually set for each channel w

Strany 252

PEL-2000 Series User Manual 84 Load D-Time Load Delay time is used to delay activating a load (up to 10 seconds) after the load key has been pre

Strany 253 - Specifications

GETTING STARTED 85 Channel Control Background External channel control is used with the Channel Control connectors. Each channel control connect

Strany 254

PEL-2000 Series User Manual 86 To determine the Percent Rating (voltage or current load input), use the following formula; VorARatingVVoltageEx

Strany 255

GETTING STARTED 87 Voltage Monitor The input voltage, like the load input current can be externally monitored with the channel control connectors.

Strany 256

PEL-2000 Series User Manual 88 Turning on the Load A load is turned on when Load On input is set to On (active low). A load is turned off when the

Strany 257

GETTING STARTED 89 Interface and File System Interface Background The PEL series support RS232, GPIB and USB remote frame control. Only one type o

Strany 258

SAFETY INSTRUCTIONS 9 Power cord for the United Kingdom When using the PEL-2002/2004 in the United Kingdom, make sure the power cord meets the follo

Strany 259

PEL-2000 Series User Manual 90 data type. Thus files are saved/recalled for each channel and each data type. Preset Data Preset data can be saved

Strany 260

GETTING STARTED 91 SEQ Data SEQ data contains Sequence data. SEQ data can only be saved to and from USB. SEQ refers to Sequence data, not Program

Strany 261 - Dimensions

PEL-2000 Series User Manual 92 Run  Active Channel (CH01~08) CHAN  Mode  Range  Static/Dynamic  CV response speed Go/NoGo  SPEC

Strany 262

GETTING STARTED 93 Save/Recall USB In order to save data from a single channel to USB, data must first be saved to the internal memory. After data

Strany 263 - EC Declaration of Conformity

PEL-2000 Series User Manual 94 To recall saved files, the reverse is not true. Files must be recalled to each channel separately. File Format S

Strany 264

GETTING STARTED 95 All Channel Directory Format 1: All Channel common directory name 2: Directory number: 0000 ~ 0099 All Channel File Forma

Strany 265 - INDEX

PEL-2000 Series User Manual 96 TUTORIALS (Step by Step Operations) Local loads ...

Strany 266

TUTORIALS 97 Local loads Local mode operation is useful to quickly test loads using the load module control panel rather than the mainframe contro

Strany 267

PEL-2000 Series User Manual 98 9. Independent control Load module Selector knobs can be configured to be independent to the mainframe. Page 188 1

Strany 268

TUTORIALS 99 Single Channel Load Single channel loads are used to manually test a DUT quickly or to configure channel settings for Program Sequenc

Komentáře k této Příručce

Žádné komentáře