GDS-2000A Programming Manual
218
String contents:
x = don’t care
1 = binary 1
0 = binary 0
:SEARCH:TRIGger:BUS:B1:CAN:DATa:SIZe 1
:SEARCH:TRIGger:BUS:B1:CAN:DATa:VALue
"01010X1X"
:SEARCH:TRIGger:BUS:B1:CAN:DATa:VALue?
>01010X1X
:SEARCH:TRIGger:BUS:B1:LIN:CONDition
Sets or returns the LIN search trigger condition.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
:SEARCH:TRIGger:BUS:B1:LIN:CONDition
{{SYNCField|IDentifier|DATA|IDANDDATA|WAKEup|S
LEEP|ERRor|?}
Parameter/
Return parameter
Sets the LIN search trigger condition
to the sync field.
Sets the LIN search trigger condition
to identifier field.
Sets the LIN search trigger condition
to the data field.
Sets the LIN search trigger condition
to identifier and data field
Sets the LIN search trigger condition
to wake up.
Sets the LIN search trigger condition
to sleep.
Sets the LIN search trigger condition
to error.