Example
:TRIGger:MODe NORMal
Sets the trigger mode to normal.
:TRIGger:HOLDoff
Description
Sets or queries the holdoff time.
Syntax
:TRIGger:HOLDoff {<NRf> | ?}
Parameter
<NRf>
Holdoff time
Return parameter
<NR3>
Returns the trigger holdoff time.
Example
:TRIGger:HOLDoff 1.00E-8
Sets the trigger holdoff time to 10ns.
:TRIGger:LEVel
Description
Sets or queries the level.
Syntax
:TRIGger:LEVel {TTL | ECL | SETTO50 | <NRf> | ?}
Related
commands
:TRIGger:TYPe
Parameter
<NRf>
Trigger level value
TTL
Sets the trigger level to TTL.
ECL
Sets the trigger level to ECL.
SETTO50
Sets the trigger level to the User level
(50% by default).
Return parameter
<NR3>
Returns the trigger level.
Example1
:TRIGger:LEVel TTL
Sets the trigger to TTL.
Example2
:TRIGger:LEVel 3.30E-1
Sets the trigger level to 330mV/mA.
:TRIGger:HLEVel
Set
Query
Set