Gw-instek GDS-2000A series Programming Manual Uživatelský manuál Strana 138

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 272
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 137
GDS-2000A Programming Manual
138
Related
commands
:TRIGger:BUS:B1:I2C:CONDition
Parameter
READ
Set read as the data direction.
WRITE
Set write as the data direction.
NOCARE
Set either as the data direction.
Return Parameter
Returns the direction (READ, WRITE, NOCARE).
Example
:TRIGger:BUS:B1:I2C:ADDRess:DIRection READ
Sets the direction to READ.
:TRIGger:BUS:B1:I2C:DATa:SIZe
Description
Sets or queries the data size in bytes for the I
2
C
bus.
Note
This setting only applies when the I
2
C trigger is set
to trigger on Data or Address/Data
Syntax
:TRIGger:BUS:B1:I2C:DATa:SIZe {<NR1> | ? }
Related
commands
:TRIGger:BUS:B1:I2C:CONDition
Parameter
<NR1>
Number of data bytes (1 to 5 ).
Return parameter
<NR1>
Returns the number of bytes.
Example
:TRIGger:BUS:B1:I2C:DATa:SIZe 3
Sets the number of bytes to 3.
:TRIGger:BUS:B1:I2C:DATa:VALue
Description
Sets or queries the triggering data value for the I
2
C
bus when the I
2
C bus is set to trigger on Data or
Address/Data.
Syntax
:TRIGger:BUS:B1:I2C:DATa:VALue {string | ? }
Related
commands
:TRIGger:BUS:B1:I2C:DATa:SIZe
Set
Query
Set
Query
Zobrazit stránku 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 271 272

Komentáře k této Příručce

Žádné komentáře