Gw-instek GSP-930 Programming Manual Uživatelský manuál Strana 145

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 195
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 144
REMOTE CONTROL
143
[:SENSe]:SEMask:BANDwidth|BWIDth
[:RESolution]
Set
Query
Description
Sets or queries the RBW for SEM measurements.
Syntax
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]
<freq>
Query Syntax
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]?
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
:SEM:BAND 2.2e+4
[:SENSe]:SEMask:BANDwidth|BWIDth
[:RESolution]:AUTO
Set
Query
Description
Turns the RBW setting to auto (on) or manual (off)
for SEM measurements or queries its state.
Syntax
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:
AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:
AUTO?
Parameter
0
Turn RBW to manual (off).
1
Turn RBW to auto (on).
OFF
Turn RBW to manual (off).
ON
Turn RBW to auto (on).
Return parameter
0
RBW is set to manual.
1
RBW is set to automatic.
Example
:SEM:BAND: AUTO OFF
[:SENSe]:SEMask:CARRier:AUTO
Set
Query
Description
Turns the PSDRef or TotalPwrRef modes to auto
(on) or manual (off) for SEM measurements or
queries their state.
Zobrazit stránku 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 194 195

Komentáře k této Příručce

Žádné komentáře