GSP-930 Programming Manual
140
<freq>
<NRf>
<NR3>
Hz
:PMET:FREQ 2e+7
[:SENSe]:PMETer:HLIMit
Set
Query
Sets or queries the power meter high limit for
pass/fail tests.
[:SENSe]:PMETer:HLIMit <ampl>
[:SENSe]:PMETer:HLIMit?
<ampl>
<NRf> power unit, default = dBm
<NR3>
Unit = current unit.
:PMET:HLIM 10
[:SENSe]:PMETer:HOLD:STATe
Set
Query
Turns the power meter Max/Min Hold function
on/off or queries its state.
[:SENSe]:PMETer:HOLD:STATe {OFF|ON|0|1}
[:SENSe]:PMETer:HOLD:STATe?
0
Turn the Max/Min Hold function off.
1
Turn the Max/Min Hold function on.
OFF
Turn the Max/Min Hold function off.
ON
Turn the Max/Min Hold function on.
0
The Max/Min Hold function is off.
1
The Max/Min Hold function is on.
:PEMT:HOLD:STAT 0
[:SENSe]:PMETer:LLIMit
Set
Query
Sets or queries the power meter low limit for
pass/fail tests.