Strany 1 - Dual Measurement Multimeter

Dual Measurement Multimeter GDM-8261A USER MANUAL GW INSTEK PART NO. 82DM-8261AEC1 ISO-9001 CERTIFIED MANUFACTURER

Strany 2

GDM-8261A User Manual 10 GETTING STARTED This chapter describes the GDM-8261A in a nutshell, including an Overview of its main features and front

Strany 3 - Table of Contents

GDM-8261A User Manual 100 Save Instrument Settings Background The GDM-8261A can save up to 5 instrument settings. The settings can save the sta

Strany 4

STORE/RECALL 101 1st display Shows the memory number 4. Select the memory number using the Up/Down key or select Del-All to delete the sav

Strany 5 - Table of Contents

GDM-8261A User Manual 102 3. Press the Down key to enter the Recall menu. 1st display Shows the memory number 4. Select the memory number

Strany 6 - SAFETY INSTRUCTIONS

SCANNER (OPTIONAL) 103 SCANNER (OPTIONAL) The optional scanner, GDM-SC1, lets you effectively measure multiple channels connected to a single

Strany 7 - Safety Guidelines

GDM-8261A User Manual 104 GDM-SC1 Scanner Specifications 2-wire channel 16 pairs Maximum current 2Arms (ch17, ch18) 4-wire channel 8 pairs Resist

Strany 8

SCANNER (OPTIONAL) 105 3. The connection terminals are revealed. Overview 16 general purpose channels are available, 8 on the left row,

Strany 9 - SAFETY INSTRUCTIONS

GDM-8261A User Manual 106 Temp. 2W RTD 2 wires (H, L) 16 (CH1 ~ 16) Temp. 4W RTD 4 wires (Input H, L + Sense H, L) 8 pairs (CH1 [input]& 9[s

Strany 10 - GETTING STARTED

SCANNER (OPTIONAL) 107 Enable scanner (Jumper J9) Set the jumper J9 on the rear side of the board accordingly. Move the jumper up (pins 3-2)

Strany 11 - GDM-8261A Characteristics

GDM-8261A User Manual 108 2. Route the wires as shown below via the two openings (left and right) at the front cover. Note When using thermoco

Strany 12 - GDM-8261A

SCANNER (OPTIONAL) 109 4. Close the top cover and tighten the screw from the bottom. Configuration Record Print out the configuration recor

Strany 13 - Measurement Keys (upper row)

GETTING STARTED 11 GDM-8261A Characteristics The GDM-8261A is a portable, dual-display digital multimeter suitable for a wide range of applic

Strany 14

GDM-8261A User Manual 110 Insert Scanner Power Off Turn the Power Off and take out the power cord. Open the GDM-8261A rear panel slot Take o

Strany 15

SCANNER (OPTIONAL) 111 Insert the scanner Insert the scanner bottom-side-up (already configured according to the procedures on page 104) into

Strany 16 - Measurement Keys (lower row)

GDM-8261A User Manual 112 Scanner Configuration Record Channel Wire color Measurement type Note CH1 H L CH2 H L CH3 H L CH4 H L CH5 H L

Strany 17

SCANNER (OPTIONAL) 113 Setup Scan Overview Scan type Simple Sets the scanned channel range, loop count, and timer length. All channels have a

Strany 18 - 250V , 2A

GDM-8261A User Manual 114 1Count# 32 4 5 6 7 8 9 10 11 12Delay DelayTimer TimerScanCH101CH 102CH 103CH 104CH101CH 102CH 103CH 104CH101CH 102CH 10

Strany 19 - 16 CHANNEL

SCANNER (OPTIONAL) 115 3. Press the Down key again. The Starting (Minimum) channel setting appears. 4. Move the cursor to the channel u

Strany 20 - Tilt Stand

GDM-8261A User Manual 116 9. Press the Enter key. The loop (step) Count setting appears. AUTOENTER 10. Move the cursor to the count number u

Strany 21 - Power Up

SCANNER (OPTIONAL) 117 4. Move the cursor to the channel using the Left/Right key, and change the value using the Up/Down key. HOLD TRIG

Strany 22

GDM-8261A User Manual 118 Range 1 ~ 999 10. Move the cursor to the count number using the Left/Right key, and change the value using the Up/

Strany 23 - BASIC MEASUREMENT

SCANNER (OPTIONAL) 119 Use External Trigger Background The GDM-8261A uses the internal trigger by default. Using an external trigger allows c

Strany 24

GDM-8261A User Manual 12 Front Panel Overview TEMP/Hz/P2/4WACV DCV ACI DCIHOLD MX/MNRELTRIGAUTO 2NDSENSORdBMATHSTORE dBmSTEPSCAN RECALLMENUREL# FI

Strany 25 - Basic Measurement Overview

GDM-8261A User Manual 120 Run Scan Overview Scan operation type Scan Measures all the specified channel ranges at each trigger event. The timer s

Strany 26 - S→M→F→S

SCANNER (OPTIONAL) 121 Recall Scan/Step Result Panel operation 1. After the Scan/Step is completed, the data is stored internally. Press the

Strany 27 - AC/DC Voltage Measurement

GDM-8261A User Manual 122 3. Press the Down key. The channel selection appears. 4. Move the cursor to the channel using the Left/Right ke

Strany 28 - Select Voltage Range

DIGITAL I/O 123 DIGITAL I/O The rear panel Digital I/O terminal outputs the result of Compare measurements to external devices. HOLDCOMP

Strany 29

GDM-8261A User Manual 124 Digital I/O Terminal Configuration Background The digital I/O terminal outputs the result of Compare measurements to c

Strany 30 - Voltage Conversion Table

DIGITAL I/O 125 Pin5 PASS signal Output. Activates when the compare result is PASS. Pin6 FAIL signal Output. Activates when the compare resu

Strany 31 - Crest Factor Table

GDM-8261A User Manual 126 2. Change the parameter using the Up/Down key. 3. Press the Enter key to confirm editing and move to the low limi

Strany 32 - AC/DC Current Measurement

DIGITAL I/O 127 Digital I/O: FAIL Out (Pin 6) and LOW Limit FAIL Out (Pin 8) are activated. Pass If the 2nd display shows Pass, the result

Strany 33 - Select Current Range

GDM-8261A User Manual 128 Application: External trigger Background The GDM-8261A uses the internal trigger by default, for example to count the f

Strany 34 - 2W/4W Resistance Measurement

DIGITAL I/O 129 2. Start trigger Press the Trig key to start triggering manually. The indicator turns On. VmSAUTOAC TRIG Reading indicator

Strany 35 - Select Resistance Range

GETTING STARTED 13 Sense HI Terminal W4WSENSEHI Accepts HI sense line in 4W resistance measurement. For details, see page 34. Input LO Termina

Strany 36 - 119.9999MΩ

GDM-8261A User Manual 130 REMOTE CONTROL Interface Overview ... 132 Configure USB In

Strany 37 - Diode Test

REMOTE CONTROL 131 TRIGger Commands ... 197 SYSTem Related Commands ...

Strany 38 - Continuity Test

GDM-8261A User Manual 132 Configure Interface Overview Interface type USB Device USB 1.1 or 2.0, TypeA, female connector. RS-232C D-sub 9 pin, m

Strany 39

REMOTE CONTROL 133 3. Press the Down key. The USB ON/OFF selection appears. 4. Press the Up/Down key to select ON or OFF. 5. Press t

Strany 40 - Select Beeper Setting

GDM-8261A User Manual 134 4. Press the Down or UP keys repeatedly to select the baud rate. 2304001152005760038400192009600 5. Press th

Strany 41 - Frequency/Period Measurement

REMOTE CONTROL 135 Configuration 1. Press the Shift key, the 2nd (Menu) key, the left key repeatedly until the TX TERM configuration menu app

Strany 42

GDM-8261A User Manual 136 Configuration 1. Press the Shift key, the 2nd (Menu) key, the left key repeatedly until the TX TERM configuration menu

Strany 43 - Temperature Measurement

REMOTE CONTROL 137 Format Description Example V Value +0.503E-4 V+U Value, Unit +0.503E-4, V DC V+C Value, Count# +0.503E-4, +00001# V+U+C Val

Strany 44

GDM-8261A User Manual 138 Insert GPIB Card Power Off Turn the Power Off and take out the power cord. Open the GDM-8261A optional communicati

Strany 45

REMOTE CONTROL 139 Power On Connect the power cord and turn On the power. Configure GPIB Interface GPIB port configuration 1. Press the

Strany 46

GDM-8261A User Manual 14 SHIFT → DCV (STEP) / EXITSHIFTDCVSTEP Starts the step measurement (page 113) using the optional scanner. ACI ACI Measures

Strany 47

GDM-8261A User Manual 140 5. To continue to the GPIB address configuration, press the Enter key. The GPIB address configuration menu appears. AU

Strany 48

REMOTE CONTROL 141 9 IFC 21 Ground (IFC) 10 SRQ 22 Ground (SRQ) 11 ATN 23 Ground (ATN) 12 SHIELD Ground 24 Single GND Insert Ethernet Card Po

Strany 49 - DUAL MEASUREMENT

GDM-8261A User Manual 142 Insert the Ethernet card Insert the Ethernet card into the slot. Close the cover by tightening the screws. Power On Con

Strany 50

REMOTE CONTROL 143 2. Press the Down key and the Left key repeatedly until LAN selection display appears. Note: The LAN menu will be selecta

Strany 51

GDM-8261A User Manual 144 Configure Ethernet Interface (RESET) Background The RESET command is used to reset the Ethernet card when new setting

Strany 52

REMOTE CONTROL 145 4. Press the Up/Down key to select YES or NO. YES will reset the Ethernet card, NO will cancel resetting the card. 5.

Strany 53 - 4W INPUT

GDM-8261A User Manual 146 2. Press the Down key and the Right key. The DHCP selection display appears. TRIG 3. Press the Down key. The DHC

Strany 54

REMOTE CONTROL 147 2. Press the Down key and the Right key twice. The IP selection display appears. Note: The IP address can only be edit

Strany 55 - ADVANCED

GDM-8261A User Manual 148 7. Repeat steps 4 to 6 for IP2, IP3 and IP4. 8. Press the Exit key to exit from the configuration menu. / EXITSHIFT

Strany 56 - Refresh Rate

REMOTE CONTROL 149 4. Use the Left/Right keys to move the cursor to the S1 value and select a digit. HOLD TRIG 5. Press the Up/Down ke

Strany 57

GETTING STARTED 15 SHIFT + Hz/P (dB) / EXITSHIFTHz/PdB Measures dB (page 59). TEMP (Temperature) TEMP Measures Temperature (page 43). SHIFT +

Strany 58 - Measure dBm/W

GDM-8261A User Manual 150 2. Press the Down key and the Right key repeatedly until the GATEWAY selection display appears. Note: The gateway can

Strany 59 - Measure dB

REMOTE CONTROL 151 8. Press the Exit key to exit from the configuration menu. / EXITSHIFT 4. DNS Configuration 1. Press the Shift key, the

Strany 60

GDM-8261A User Manual 152 5. Press the Up/Down key to edit the selected digit. 6. Press the Enter key to confirm and automatically go o

Strany 61 - Max/Min Measurement

REMOTE CONTROL 153 3. Press the Down key. The MAC address appears. 4. Press the Exit key to exit from the configuration menu. / EXITS

Strany 62 - Relative Value Measurement

GDM-8261A User Manual 154 3. Press the Down key. The Telnet port appears. 4. Change the telnet port using the Left/Right and Up/Down keys. HO

Strany 63

REMOTE CONTROL 155 Return to Initial Settings 1. Press the Shift key, the 2nd (Menu) key, and the Left key repeatedly until the SET LAN con

Strany 64 - Hold Measurement

GDM-8261A User Manual 156 Note If the GDM-8261A is returned to the initial settings, a reset is performed automatically (page 144) after exitin

Strany 65 - Compare Measurement

REMOTE CONTROL 157 3. Press the Down key. The WEB PW ON/OFF selection appears. 4. Press the Up/Down key to select ON or OFF. When set

Strany 66

GDM-8261A User Manual 158 3. Run this query via the terminal application: *idn? The command will return the instrument manufacturer, model numbe

Strany 67

REMOTE CONTROL 159 Web Control Interface The web control interface is accessible with the optional Ethernet card. The web control interface al

Strany 68 - Math Measurement

GDM-8261A User Manual 16 Measurement Keys (lower row) AUTO/ENTER AUTOENTER As the AUTO key, selects the measurement range automatically. As the E

Strany 69

GDM-8261A User Manual 160 Note If the password dialog box or the Welcome Page fail to appear when WEB PW is set to ON, please ensure JavaScript

Strany 70 - Measure Percentage

REMOTE CONTROL 161 3. View and Modify LAN Configuration The current Ethernet settings can be viewed and modified from the web control interfac

Strany 71

GDM-8261A User Manual 162 3. The View & Modify Configuration page allows you to:  View the instrument name, firmware revision of the E

Strany 72

REMOTE CONTROL 163 Command Syntax The commands are partially compatible with IEEE488.2 (1992) and SCPI (1994) standard. Commands are NON-case

Strany 73 - SDC AUTO

GDM-8261A User Manual 164 Message Separator EOL or , Command separator, user configurable (excluding GPIB), see page 135. Square Brackets [ ]

Strany 74 - SYSTEM/DISPLAY

REMOTE CONTROL 165 CONFigure2:FUNCtion? ... 176 CONFigure2:RANGe? ...

Strany 75 - View Serial Number

GDM-8261A User Manual 166 [SENSe:]TEMPerature:FRTD:TYPE? ... 182 [SENSe:]TEMPerature:FRTD:AL

Strany 76 - Trigger Setting

REMOTE CONTROL 167 [SENSe:]DIGital:SHIFt? ... 186 [SENSe:]UNIT...

Strany 77

GDM-8261A User Manual 168 [SENSe:]VOLTage:DC:RESolution? ... 190 [SENSe:]VOLTage:AC:RESolut

Strany 78 - Set Trigger Delay

REMOTE CONTROL 169 CALCulate:REL:REFerence? ... 195 CALCulate:LIMit:LOWer ..

Strany 79 - Filter Setting

GETTING STARTED 17 SHIFT → REL (RELative base) / EXITSHIFTRELREL# Manually sets the reference value for the Relative value measurement (page 6

Strany 80 - Digital Filter Setting

GDM-8261A User Manual 170 TRIGger:COUNt? ... 199 TRACe:DATA? ...

Strany 81

REMOTE CONTROL 171 *OPC ... 204 *PSC?...

Strany 82 - Analog Filter Settings

GDM-8261A User Manual 172 CONFigure Commands CONFigure:VOLTage:DC Sets measurement to DC Voltage on the first display and specifies range/resolut

Strany 83 - Display Setting

REMOTE CONTROL 173 CONFigure:FRESistance Sets measurement to 4W Resistance on the first display and specifies the range/resolution. Parameter:

Strany 84 - D-Shift Settings

GDM-8261A User Manual 174 CONFigure:TEMPerature:RTD Sets the measurement mode to 2W RTD measurement mode on the first display. Sets the sensor typ

Strany 85 - Input Resistance Setting

REMOTE CONTROL 175 CONFigure2:VOLTage:AC Sets measurement to AC Voltage on the second display and specifies range/resolution. Parameter: [None

Strany 86 - AC Bandwidth Setting

GDM-8261A User Manual 176 CONFigure2:FREQuency Sets measurement to Frequency on the second display and specifies range/resolution. Parameter: [Non

Strany 87

REMOTE CONTROL 177 Measure Commands MEASure:VOLTage:DC? Returns the DC voltage measurement on the first display. Parameter: [None] | [Range(&

Strany 88 - ADC Setting

GDM-8261A User Manual 178 MEASure:FRESistance? Returns the 4W resistance measurement on the first display. Parameter: [None] | [Range(<NRf>

Strany 89

REMOTE CONTROL 179 MEASure:TEMPerature:FRTD? Returns the 4W RTD temperature for the selected sensor type on the first display. Parameter:[NONE

Strany 90 - Auto-Gain

GDM-8261A User Manual 18 Rear Panel Overview RS232CAL KEYDIGITAL I / OSER.NO. LABEL50 , 60 , 400Hz25VA MAX.LINE RATINGFUSE LINE100VAC120VAC220

Strany 91

GDM-8261A User Manual 180 MEASure2:RESistance? Returns the 2W resistance measurement on the second display. Parameter: [None] | [Range(<NRf>

Strany 92 - ADC Speed Setting

REMOTE CONTROL 181 [SENSe:]TEMPerature:RJUNction:SIMulated Set temperature simulation value. Parameter: <NRf>(0.00 ~ 50.00) Example: SEN

Strany 93 - Frequency / Period Settings

GDM-8261A User Manual 182 [SENSe:]TEMPerature:FRTD:TYPE Sets the 4W RTD sensor type. Parameter: Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)

Strany 94

REMOTE CONTROL 183 [SENSe:]AVERage:TCONtrol Selects the digital filter. Parameter: MOV | REP Example: SENS:AVER:TCON MOV Sets the digital filt

Strany 95 - Gate Time Setting

GDM-8261A User Manual 184 [SENSe:]FILTer:STATe Turns the analog filter On/Off. Parameter: ON | OFF Example: SENS:FILT:STAT ON Turns the analog f

Strany 96 - Identification Settings

REMOTE CONTROL 185 [SENSe:]PERiod:INPutjack? Returns the assigned input port used for the period function. Return Parameter: VOLT, 1A, 10A [SE

Strany 97 - STORE/RECALL

GDM-8261A User Manual 186 [SENSe:]CURRent:DETect? Returns the auto-detect status for the current functions. Return Parameter: 0|1 1=ON, 0=OFF [

Strany 98 - Store Measurement Record

REMOTE CONTROL 187 [SENSe:]VOLTage:DC:RANGe? Returns the DC Voltage measurement range. Parameter: [None] | [MIN | MAX] [SENSe:]VOLTage:AC:RANG

Strany 99

GDM-8261A User Manual 188 [SENSe:]FRESistance:RANGe Sets the 4W resistance measurement range. Parameter: Range(<NRf> | MIN | MAX) Example:

Strany 100 - Save Instrument Settings

REMOTE CONTROL 189 [SENSe:]VOLTage:AC:RANGe:AUTO? Returns the AC voltage Auto-range settings. Return parameter: 0|1, 0=OFF, 1=ON [SENSe:]CURRe

Strany 101 - Recall Instrument Settings

GETTING STARTED 19 USB device port Accepts a USB device cable for remote control; Type A, female connector. For remote control details, see p

Strany 102 - / EXITSHIFT

GDM-8261A User Manual 190 [SENSe:]FREQuency:VOLTage:RANGe:AUTO Sets the Frequency Auto-range settings on/off. Parameter: ON|OFF Example: SENS:FREQ

Strany 103 - SCANNER (OPTIONAL)

REMOTE CONTROL 191 [SENSe:]CURRent:DC:RESolution? Returns the DC Current resolution. Parameter: [None] | [MIN | MAX] [SENSe:]CURRent:AC:RESolu

Strany 104 - Scanner Installation

GDM-8261A User Manual 192 [SENSe:]DIODe:RESolution Sets the Diode measurement resolution. The resolution depends on the rate and range settings. P

Strany 105

REMOTE CONTROL 193 [SENSe:]VOLTage:DC:NPLCycles Sets the integration time for DC Voltage measurements in PLCs (power line cycles). Where one P

Strany 106

GDM-8261A User Manual 194 [SENSe:]FRESistance:NPLCycles Sets the integration time for 4W resistance measurements in PLCs (power line cycles). Wher

Strany 107 - Connect Wires

REMOTE CONTROL 195 CALCulate:HOLD:REFerence? Returns the percentage threshold from the Hold function. CALCulate:REL:REFerence Sets the referen

Strany 108

GDM-8261A User Manual 196 CALCulate:STORe:COUNt Set the number of measurement counts that are recorded with the Store measurement function. Para

Strany 109 - SCANNER (OPTIONAL)

REMOTE CONTROL 197 CALCulate:MATH:MMFactor Sets the scale factor M for math measurements. Para meter: <NRf> | MIN | MAX Example: CALC:MA

Strany 110 - Insert Scanner

GDM-8261A User Manual 198 VAL1? Returns the 1st display reading in the unit format specified in the Configuration menu (Return Format, page 136) o

Strany 111

REMOTE CONTROL 199 TRIGger:AUTO? Returns the Trigger Auto mode. Return parameter: 0|1, 0=OFF, 1=ON SAMPle:COUNt Sets the number of samples.

Strany 112 - Scanner Configuration Record

This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied,

Strany 113 - Setup Scan

GDM-8261A User Manual 20 Set Up Tilt Stand Tilt stand steps Pull out the handle sideways and rotate it. Place the unit horizontally, Or in

Strany 114 - Setup Simple Scan

GDM-8261A User Manual 200 SYSTem Related Commands SYSTem:BEEPer:STATe Selects the beeper mode; no beep, beep on fail and beep on pass. Parame

Strany 115 - HOLD TRIG

REMOTE CONTROL 201 SYSTem:OUTPut:FORMat? Returns the output format. Return parameter: (0|1|2|3) (0=V, 1=V+U, 2=V+C, 3=V+U+C) SYSTem:OUTPut:EOF

Strany 116 - Setup Advanced Scan

GDM-8261A User Manual 202 SYSTem:SCPi:MODE Sets the SCPI mode. The SCPI mode is used to determine whether the *IDN? query returns the “Normal” or

Strany 117

REMOTE CONTROL 203 STATus Report Commands STATus:QUEStionable:ENABle Set bits in the Questionable Data Enable register. STATus:QUEStionable:E

Strany 118

GDM-8261A User Manual 204 *ESR? Returns SESR (Standard Event Status Register) contents. Example: *ESR? >198 Returns 198. SESR=1100011

Strany 119 - Use External Trigger

REMOTE CONTROL 205 ROUTe Commands ROUTe:CLOSe Close a specified scanner channel. Parameter: <NR1>(101~118) Example: ROUT:CLOS 102 Clo

Strany 120 - Run Scan

GDM-8261A User Manual 206 ROUTe:CHANnel Advanced configuration mode for the scanner channels. The channel function, voltage and Auto-range mode ca

Strany 121

REMOTE CONTROL 207 ROUTe:ADVance Turns the scanner Advanced mode on/off. Parameter: ON|OFF Example: ROUT:ADV OFF Turns advanced scanner mode o

Strany 122

GDM-8261A User Manual 208 FETCh? Transfer the stored readings to the output buffer. DATA:POINts? Returns the number of readings. Parameter: None |

Strany 123 - DIGITAL I/O

FAQ 209 FAQ ● I pressed the EXIT key but cannot get out of Scanner mode. ● The GDM-8261A performance does not match the specifications. I p

Strany 124

GETTING STARTED 21 Power Up Steps 1. Ensure the correct line voltage is lined up with the arrow on the fuse holder. If not, see page 213 to

Strany 125 - >10μs (R=1kΩ)

GDM-8261A User Manual 210 2. SYST: DISP OFF 3. SYST: OUTP:FORM 0 4. TRIG:DEL 0 5. SENS:AVER:STAT OFF 6. SAMP:COUN 2400 7. VAL1 ? If the

Strany 126

APPENDIX 211 APPENDIX System Info Firmware Version ...212 Fuse Replacement Replace AC Source F

Strany 127

GDM-8261A User Manual 212 Firmware Version Background Firmware version is available from the system menu. Firmware version Shows the GDM-8261A f

Strany 128

APPENDIX 213 Fuse Replacement Replace AC Source Fuse Steps 1. Take off the power cord and remove the fuse socket using a minus driver. 1 0 0

Strany 129

GDM-8261A User Manual 214 Replace Input Current Fuse Step 1. Press the Fuse holder. 2. The fuse holder comes out. Replace the fuse inserted

Strany 130 - REMOTE CONTROL

APPENDIX 215 Menu Tree Menu Tree Background The menu tree diagram shown on the next page represents the configuration menu that is accessed b

Strany 131 - 131

GDM-8261A User Manual 216 Configuration Menu Tree 2NDMENUMEAS SET ADC HZ/P I/OCONT D-SHIFT INPUT R AC BW I-DETSPEED A-GAING-TIMERUSB GPIBVER LIGH

Strany 132 - Configure Interface

APPENDIX 217 Status system The diagram below is a description of the status system For the following command sets, please refer to the di

Strany 133 - Configure RS-232C Interface

GDM-8261A User Manual 218 Specifications General Note  All specifications are ensured only under a single display.  At least 1 hour of warm-

Strany 134 - Set the EOL Character

APPENDIX 219 DC Characteristics [3] DC Voltage [1] Range [4] 24 Hour 23˚C±1˚C 90 Day 23˚C±5˚C 1 Year 23˚C±5˚C Temperature Coefficient/˚C [6]

Strany 135 - Set the Separation Character

GDM-8261A User Manual 22 7. Then the default setting appears. Example: DCV, Auto, 100mV range VmSDCAUTO

Strany 136 - Set the Return Format

GDM-8261A User Manual 220 Measuring Characteristics DC Voltage Input Resistance Range 0.1V 10MΩ or >10GΩ Selectable 1V 10MΩ or >10GΩ S

Strany 137

APPENDIX 221 function. Without REL function, add 0.2 Ω additional error in 2-wire ohms function. [ 6 ] 0˚C~18˚C, 28˚C~55˚C [ 7 ] Accuracy sp

Strany 138 - Insert GPIB Card

GDM-8261A User Manual 222 10.00000 mA 3Hz – 5Hz 1.1 + 0.06 1.1 + 0.06 1.1+0.06 0.2+0.006 5Hz – 10Hz 0.35 + 0.06 0.35 + 0.06 0.35+0.06 0.1+0.006

Strany 139 - Configure GPIB Interface

APPENDIX 223 Measuring Characteristics True RMS AC Voltage Measurement Method: AC-coupled True RMS – measures the ac component of input with

Strany 140

GDM-8261A User Manual 224 [ 10 ] All speeds need Fixed range and Trigger Delay=0. Frequency and Period Characteristics Frequency Period [3] [7]

Strany 141 - Insert Ethernet Card

APPENDIX 225 [ 6 ] Need Fixed ACI/ACV range and Trigger Delay=0. [ 7 ] To meet the specifications accuracy, Slow rate setting is needed. Tempe

Strany 142

GDM-8261A User Manual 226 Dimensions 22088220264.4350.310788309.18.416325.176.815.75710.7201.951019 7019 19All dimensions are shown in millimeter

Strany 143

APPENDIX 227 EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. (1) No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City, Taiwan

Strany 144

GDM-8261A User Manual 228 INDEX A AC bandwidth setting ... 86 ADC speed setting ...

Strany 145 - 1. DHCP

INDEX 229 Gateway ... 149 Initial settings ... 154 IP 1

Strany 146 - Configure Ethernet IP

BASIC MEASUREMENT 23 BASIC MEASUREMENT ACVDCVACIDCI2/4W/Hz/PTEMP Overview Basic Measurement Overview ... 25 Ref

Strany 147

GDM-8261A User Manual 230 triggering ... 119 Scanner configuration record/log ... 112

Strany 148

GDM-8261A User Manual 24 Temperature Temperature Measurement ... 43 Select Thermocouple Type ...

Strany 149

BASIC MEASUREMENT 25 Basic Measurement Overview Background Basic measurement refers to the eight types of measurements assigned to the upper

Strany 150

GDM-8261A User Manual 26 Refresh Rate Function S M F (Readings/s) Continuity / Diode 100 200 300 DCV/DCI/100Ω~ 100MΩ (Accurate) 5 60 240 DCV

Strany 151

BASIC MEASUREMENT 27 Manual/Automatic Triggering Automatic triggering (default) The GDM-8261A triggers according to the refresh rate. See the

Strany 152 - View MAC Address

GDM-8261A User Manual 28 Select Voltage Range Auto range To turn the automatic range selection On/Off, press the AUTO key. AUTO Manual range Pres

Strany 153 - Configure Telnet Port

BASIC MEASUREMENT 29 The DC voltage range should be manually selected when any of the following conditions are true: 1. When DCV measureme

Strany 154 - Return to Initial Settings

Table of Contents 3 Table of Contents SAFETY INSTRUCTIONS ... 6 Safety Symbols ...

Strany 155

GDM-8261A User Manual 30 Voltage Conversion Table This table shows the relationship between AC and DC reading in various waveforms. Waveform Peak

Strany 156 - View Web Password Settings

BASIC MEASUREMENT 31 Crest Factor Table Background Crest factor is the ratio of the peak signal amplitude to the RMS value of the signal. It

Strany 157

GDM-8261A User Manual 32 AC/DC Current Measurement Background The GDM-8261A has two input ports for current measurement. A 1A terminal for curren

Strany 158 - GDM-8261A User Manual

BASIC MEASUREMENT 33 Select Current Range Auto range To turn the automatic range selection On/Off, press the AUTO key. AUTO Manual range Pres

Strany 159 - Web Control Interface

GDM-8261A User Manual 34 The DC current range should be manually selected when the following conditions are true: 1. When DCI measurement is u

Strany 160

BASIC MEASUREMENT 35 2. 2W/4W resistance mode display appears WS2WkAUTO 2W or 4W + Ω Indicates 2W or 4W Resistance mode AUTO Indicates Autom

Strany 161

GDM-8261A User Manual 36 10MΩ 10Ω 11.99999MΩ 100MΩ 100Ω 119.9999MΩ Note For more detailed range, see the specifications at page 217.

Strany 162

BASIC MEASUREMENT 37 Diode Test Background Diode test checks the forward bias characteristics of a diode by running a constant forward bias

Strany 163 - 1 2 3 4 5

GDM-8261A User Manual 38 Continuity Test Background Continuity test checks that the resistance in the DUT is low enough to be considered continuo

Strany 164 - Command Set

BASIC MEASUREMENT 39 Set Continuity Threshold Background Continuity threshold defines the maximum resistance allowed in the DUT when testing

Strany 165 - 165

GDM-8261A User Manual 4 View Serial Number ... 75 Trigger Setting ...

Strany 166

GDM-8261A User Manual 40 Select Beeper Setting Background Beeper setting defines how the GDM-8261A notifies the continuity test result to the use

Strany 167 - 167

BASIC MEASUREMENT 41 Frequency/Period Measurement 1. Activate frequency/period measurement To measure Frequency, press the Hz/P key once. Hz/

Strany 168

GDM-8261A User Manual 42 Select Frequency/Period Voltage Range Frequency/Period mode To select between period/frequency voltage range, press the

Strany 169 - 169

BASIC MEASUREMENT 43 Temperature Measurement Background The GDM-8261A can measure temperature using either thermocouples or RTD sensors. For

Strany 170

GDM-8261A User Manual 44 Select Thermocouple Type Background The GDM-8261A accepts thermocouple inputs and calculates the temperature from the vo

Strany 171 - 171

BASIC MEASUREMENT 45        5. Confirm and go back to the default display Press the Enter key to confirm. Press the Exit key to go ba

Strany 172 - CONFigure Commands

GDM-8261A User Manual 46 2. Edit reference temperature Use the Left/Right key to move the cursor, and use the Up/Down key to change the value. Def

Strany 173 - Ω at the maximum

BASIC MEASUREMENT 47 3. Select sensor Press the Up and Down keys to highlight the RTD sensor type. RTD Type: PT 100, PT 3916, PT 385, F

Strany 174

GDM-8261A User Manual 48 Press the Down key twice. The RTD selection menu appears on the display. Use the Up/Down keys to select USER. 3.

Strany 175 - CONF2:RES 10e3,MIN

DUAL MEASUREMENT 49 DUAL MEASUREMENT Dual Measurement Background The dual measurement mode allows you to use the 2nd display to show anothe

Strany 176

Table of Contents 5 INDEX ... 228

Strany 177 - Measure Commands

GDM-8261A User Manual 50 Amplifier Output INPUTV WCOMVVoTAMPVoAC NoiseDC offset DC Power Supply Output DC outputDC Power SupplyINPUTV WCOMVAC ri

Strany 178

DUAL MEASUREMENT 51 ACV Hz  Measure the frequency response of devices such as amplifiers or buffers*. * The frequencies of the amplifier ou

Strany 179

GDM-8261A User Manual 52 2nd Measurement item setting Press the 2nd key, then the target item (example: ACV). The display updates the measurement

Strany 180 - SENSe Commands

DUAL MEASUREMENT 53 Connect the test leads and measure When using the dual measurement function, the connection method and number of test lea

Strany 181 - 181

GDM-8261A User Manual 54 Voltage/Frequency/Period and Current Measurement INPUTV WCOM0~10A0~1.2ALoad Note: DC Current measurements will be displ

Strany 182

ADVANCED MEASUREMENT 55 ADVANCED MEASUREMENT /dBmHz/PdBMX/MNFILTERRELREL#HOLDCOMPTRIGINT/EXT2/4WMATH Overview Advanced Measurement Overview

Strany 183 - 183

GDM-8261A User Manual 56 Advanced Measurement Overview Background Advanced measurement mainly refers to the type of measurement which uses the r

Strany 184

ADVANCED MEASUREMENT 57 Refresh Rate Function S M F (Readings/s) Continuity / Diode 100 200 300 DCV/DCI/100Ω~ 100MΩ (Accurate) 5 60 240 DC

Strany 185 - 185

GDM-8261A User Manual 58 Manual triggering Press the Trig key to trigger the measurement manually. The trigger must be set to external (EXT) for m

Strany 186

ADVANCED MEASUREMENT 59 Select reference resistance To change the reference resistance, press the Up/Down key. The new resistance appears in t

Strany 187 - 187

GDM-8261A User Manual 6 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow when operating the GDM-8261A

Strany 188

GDM-8261A User Manual 60 dB result appears dBSDCVm dB Indicates dB measurement -00.617mV Indicates the present Voltage reading dBmref Press the

Strany 189 - 189

ADVANCED MEASUREMENT 61 Max/Min Measurement Applicable to ACVDCVACIDCI2/4WHz/PTEMP Background Maximum and Minimum measurement stores the high

Strany 190

GDM-8261A User Manual 62 Relative Value Measurement Applicable to ACVDCVACIDCI2/4WHz/PTEMP Background Relative measurement stores a value, typica

Strany 191 - 191

ADVANCED MEASUREMENT 63 V REL Indicates Relative measurement 1st display Shows the reference value (to full scale) 2nd display Indicate

Strany 192

GDM-8261A User Manual 64 Hold Measurement Applicable to ACVDCVACIDCI2/4WHz/PTEMP Background The Hold Measurement function retains the current mea

Strany 193 - 193

ADVANCED MEASUREMENT 65 Compare Measurement Applicable to ACVDCVACIDCI2/4WHz/PTEMP Background Compare measurement checks and updates if the

Strany 194 - CALCulate Commands

GDM-8261A User Manual 66 3. Low limit setting V 1st display Shows the low limit value 2nd display Indicates low limit setting 1. Use the Left

Strany 195 - 195

ADVANCED MEASUREMENT 67 5. Result High If the 2nd display shows High, the result is above the High limit. Digital I/O: FAIL Out (Pin 6) a

Strany 196

GDM-8261A User Manual 68 Math Measurement Applicable to ACVDCVACIDCI2/4WHz/PTEMP Background Math measurement runs four types of mathematical ope

Strany 197 - TRIGger Commands

ADVANCED MEASUREMENT 69 2. Change the parameter using the Up/Down key. 3. Press the Enter key to confirm editing and move to offset setti

Strany 198

SAFETY INSTRUCTIONS 7 Safety Guidelines General Guideline CAUTION  Make sure that the voltage input level does not exceed DC1000V/AC750V.

Strany 199 - 199

GDM-8261A User Manual 70 Measure 1/X 1. Activate 1/X Press the Shift key, the 2/4W (Math) key, the Down key twice. The 1/X setting appears. / EXI

Strany 200 - SYSTem Related Commands

ADVANCED MEASUREMENT 71 1. Use the Left/Right key to move the cursor (flashing point) between the digits and decimal point. HOLD TRIG 2.

Strany 201 - 201

GDM-8261A User Manual 72 2. Set Count Press the Enter key to set the number of measurements (counts) that will be used for the Stats function.

Strany 202

ADVANCED MEASUREMENT 73 3. Press the Enter key to confirm editing and to start measurement. AUTOENTER 3. View Data SDC AUTOMATH 1st displ

Strany 203 - IEEE 488.2 Common Commands

GDM-8261A User Manual 74 SYSTEM/DISPLAY CONFIGURATION ACVDCVACIDCIAUTORATEMX/MNFILTER2NDMENU Refresh Rate Refresh Rate Setting ...

Strany 204

SYSTEM/DISPLAY CONFIGURATION 75 Refresh Rate Setting Background Refresh rate defines how frequently the GDM-8261A captures and updates the me

Strany 205 - ROUTe Commands

GDM-8261A User Manual 76 2. Press the Down key. The serial number is shown on the display. 1st display Shows 2 characters (AA~ZZ). 2nd

Strany 206 - 101,VOLT,0.1,ON

SYSTEM/DISPLAY CONFIGURATION 77 DB-9, female Digital I/O pin assignment 6FAIL Out7 8 912 3 4 5High Limit FAIL OutVCC OutNCPASS OutExternal

Strany 207 - 207

GDM-8261A User Manual 78 Set Trigger Delay Background Trigger delay defines the time delay between triggering and measurement start. The default

Strany 208

SYSTEM/DISPLAY CONFIGURATION 79 Filter Setting Digital Filter Overview Filter basics The GDM-8261A’s internal digital filter converts the ana

Strany 209 - FAQ

GDM-8261A User Manual 8 Fuse WARNING  Fuse type: 0.315AT 100/120VAC 0.125AT 220/240 VAC  Make sure the correct type of fuse is in

Strany 210

GDM-8261A User Manual 80 Filter count Filter count defines the number of samples to be averaged per reading. More samples offer low noise but a l

Strany 211 - APPENDIX

SYSTEM/DISPLAY CONFIGURATION 81 2. Select the filter type using the Up/Down key.   3. Move the cursor to filter count using the Left/R

Strany 212 - Firmware Version

GDM-8261A User Manual 82 Analog Filter Settings Background The analog filter is a single order low pass filter that can be turned on to attenuate

Strany 213 - Fuse Replacement

SYSTEM/DISPLAY CONFIGURATION 83 6. Press the Enter key to confirm your selection. Press the Exit key to go back to the default display. AUTO

Strany 214 - Replace Input Current Fuse

GDM-8261A User Manual 84 4. Select the level using the Up/Down key. 5. Press the Enter key to confirm your selection. Press the Exit key to g

Strany 215 - Menu Tree

SYSTEM/DISPLAY CONFIGURATION 85 1st display Shows the D-Shift setting 4. Select the setting using the Up/Down keys. 5. Press the Ente

Strany 216 - SHIFT/ EXIT

GDM-8261A User Manual 86 1st display Shows the input resistance setting 4. Select the setting using the Up/Down keys. 5. Press the Enter

Strany 217 - Status system

SYSTEM/DISPLAY CONFIGURATION 87 3. Press the Down key. The input bandwidth setting appears. 1st display Shows the bandwidth setting 4.

Strany 218 - Specifications

GDM-8261A User Manual 88 2. Press the Right key, then the down key. Press the left key. The Current detect menu appears. TRIG HOLD 3. Pres

Strany 219 - DC Characteristics

SYSTEM/DISPLAY CONFIGURATION 89 Buffer, A/D driver and ADC vary over time, and thus the total offset will also vary over time. Auto Zero ded

Strany 220

SAFETY INSTRUCTIONS 9 Power cord for the United Kingdom When using the GDM-8261A in the United Kingdom, make sure the power cord meets the fol

Strany 221 - AC Characteristics

GDM-8261A User Manual 90 1st display Shows A-Zero setting 3. Select the setting using the Up/Down key. 4. Press the Enter key to confirm

Strany 222

SYSTEM/DISPLAY CONFIGURATION 91 2. Press the Down key and then the Left key twice to choose A-GAIN. Press the Down key. The A-GAIN setting a

Strany 223 - 223

GDM-8261A User Manual 92 ADC Speed Setting Background The analog to digital converters have a Quick and Accurate Speed setting. The ADC Speed set

Strany 224

SYSTEM/DISPLAY CONFIGURATION 93 4. Use the Up/Down keys to select either ACCUR or QUICK. 5. Press the Enter key to confirm your selection

Strany 225 - Slow rate

GDM-8261A User Manual 94 4. Press the Enter key to confirm your selection. Press the Exit key to go back to the default display. AUTOENTER/ EXIT

Strany 226 - Dimensions

SYSTEM/DISPLAY CONFIGURATION 95 Gate Time Setting Background The gate time settings determine the accuracy of the frequency and period measu

Strany 227 - EC Declaration of Conformity

GDM-8261A User Manual 96 Identification Settings Changing the Identification String Background The *IDN? query returns the manufacturer, model nu

Strany 228

STORE/RECALL 97 STORE/RECALL The GDM-8261A can store and recall measurement history (for up to 9999 counts) as well as the instrument settings

Strany 229 - 229

GDM-8261A User Manual 98 Store Measurement Record Background The GDM-8261A can log up to 9999 measurement results (counts) which can be stored an

Strany 230

STORE/RECALL 99 Recall Measurement Record Background The GDM-8261A can recall previously recorded measurement results for observation and ana